USING
THE
GRCM
a
Cleaning the
GRCM
,
7.
Purge the ASW Tray Delivery line:
a. On the ASW tray, turn off the solvent switch and turn on the purge switch.
b.
Turn Control Valve
1
to SOLVENT for
1
minute.
c. Turn off the purge switch on the ASW tray.
8.
Purge the Gas Delivery Manifold:
a. Turn Control Valve
1
to
DRAIN
GAS for
2
to
5
minutes to remove all solvent
from the
Gas
Delivery Manifold.
b.
Turn Control Valve
1
to CLOSED.
9.
Detach Gas Delivery Manifold Line from the Quest
210
and reattach to
outlet on the GRCM control box.
10.
Reattach Upper Manifold Delivery Line to the Quest
210.
FEMALE LUERS
ON
THE
QUEST LOWER MANIFOLD
1.
Wash the hers with solvent using a squeeze bottle
2.
Wipe them clean with cotton tips as used used to clean the Quest.