Operating the Quest
210
Starting a Reaction
Temperature Control
-
The allowable reaction temperature range on the Quest is
-40
to
+130.
For a particular
reaction, do not program a temperature set-point which exceeds the boiling point of the
reaction solvent.
Each side of the Quest may have a different set point. To achieve sub-ambient
temperatures, an external recirculating chiller is required. When using the chiller, only
connect it to one side of the Quest at a time.
WARNING When heating, the Reactor
Assembly will become hot. The symbol at the
right is to warn of hot surface. Use caution when
working near or on the Reactor Assembly.
Allow it to cool before attempting to access RVs.
Prior to starting a heating step, set both Manifold Control Valves to "Closed". This
prevents solvent loss during heating:
drain
solvent
G
closed
utility
1
metered gas
vent
0
closed
utility
2
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com