Agitation Control Screen
Agitation Control Screen
MixEvery
:
2.0
sec
UpStroke
:
1.0
sec
%
Upward
:
50
%
I--
Not Agitating
-
-
-
The Agitation Control screen allows you to set the agitation parameters.
The Agitation Control screen contains three parameters: MixEverysec, UpStroke
,
and
%
Upward. The MixEverysec parameter is the agitation frequency. That is, the
elapsed time between consecutive upstrokes.
Use the PARAMETER SETTING buttons to increase or decrease the MixEverysec
parameter.
The UpStroke parameter is the amount of time the agitation bar will be in the upper
position (This number is necessarily less than the agitation frequency value)
The
%
Upward parameter controls the upstroke in the same manner as the "Upstroke"
parameter, but is expressed in terms of a percentage of the agitation frequency, rather
than a time.
If you change the UpStroke parameter, then the
%
Upward parameter will automati-
cally change to correspond to the UpStroke parameter. Likewise, if you change the
%
Upward parameter, then the Upstroke parameter will change automatically to corre-
spond to the
%
Upward parameter.
For example, if the MixEvery parameter is set to 2.0 seconds and the
%
Upward
parameter to 70%, the UpStroke parameter on the screen automatically changes to
1.4
seconds.