Collection of Liquid Reaction Products
1.
Toggle the "Upper Manifold Membrane" switch to "Open."
2.
Prior to product collection, it is recommended that the system be vented to relieve
any residual pressure that can cause splashing during product collection. To vent the
vessels, rotate the Manifold Control Valves as shown below. Slowly approach the
"Vent" setting on the valve to vent the pressure slowly to avoiding "bumping" the
solution.
drain
solvent
n
closed
utility
1
metered gas
vent
0
closed
utility
2
3.
Carefully lower and remove the Glass Collection Tank, and drain its contents to an
appropriate storage container.
4.
Place the desired collection rack and vials on the waste collection tray.
5.
Set the Manifold Control Valves as follows to pressurize the head space over the
reaction vessels:
drain
solvent closed
utility
1
metered
gas
vent
0
closed
u
utility
2
6.
Open each drain valve on the lower manifold (down lever position) one at a time.
The solution should drain into the collection vial. Return the valve to closed (up
position) after each collection. The flow rate of the Metered gas may need to be
adjusted for the this step