Figure
7.
Remove the
RV
from the holder
With your fingers, remove the RV from the Reactor Unit (Figure
7)
Using a sharp pair of scissors, cut the top of the RV off below the crimp from the RV
extraction tool.
Use the external hand magnet provided in the Quest starter kit to remove the magnet
mixers from the RV. Remove the magnets by sliding the magnets out through the top
of the RV.
Wash the magnetic mixers with an appropriate solvent such as acetone.
Remove the Luer Plugs (twist and pull) from the Upper Manifold, rinse them with
an
appropriate solvent (e.g., acetone), and wipe clean.
Use a wash bottle or syringe to rinse the Luer Ports of the Upper Manifold with an
appropriate solvent. Hold a soft cloth or paper towel beneath the Upper Manifold to
absorb the rinse solvent.
Use a cotton swab wetted with an appropriate solvent to wipe the inside of the Luer
Ports of the Upper Manifold and to wipe around the Upper Manifold male Teflon RV
fittings.
-
10.
Use a cotton swab wetted with
an
appropriate solvent to wipe clean the male Teflon
RV fittings of the Lower Drain Manifold. Lift the RV collars and wipe around the
bottom of the male Teflon RV fittings.
1
1.
Finally, re-examine the Upper and Lower Manifold male RV fittings. Clean the RV
fittings and remove any residue or particulates with a soft cloth
and
an appropriate
solvent. Do not allow any particulates to enter the drain holes of the Lower Manifold
male RV fittings. Take care not to scratch the male
RV fittings as this may result in
leaks.
Argonaut Technologies
Part Number 102298
04/02/98