Raise the handle so the Lower Manifold Drain Valves are closed.
Add
-4
mL DCM
to the RVs by turning the Manifold Control Valves to
"SOLVENT" and "VENT". When the delivery is complete, purge the Upper
Manifold by
slowly
turning the
left
Manifold Control Valve to "DRAIN GAS".
Close the Upper Manifold by turning both of the Manifold Control Valves to
"CLOSED".
Turn the agitation on by pushing the Agitation ON soft key located to the right side of
the Quest Controller Unit display.
Agitate the resin for
-5
minutes.
Repeat steps 1-8 two more times using DCM.
10. ~r&n the
RVS
as described in steps 1-4.
G.
Addition of Anhydrous NMP Under a
N2
Sweep
via
the Upper Manifold
Luer Ports
Purge the resin for -30 seconds with nitrogen
a.
Turn the Manifold Control Valves to "DRAIN GAS" and "CLOSED".
b. Toggle the
10
Lower Manifold Drain Valves open.
Close the Lower Manifold Drain Valves and vent the Upper Manifold by turning the
Manifold Control Valves to "CLOSED" and "VENT".
Turn the Manifold Control Valves to "UTILITY 1" and "METERED GAS". This
provides
a
low flow nitrogen sweep of the RVs during the addition of anhydrous
NMP. Metered Gas will flow out of the Utility 1 port to the oil bubbler.
Drain Gas Metered Gas
Utility 1 Utility
2
Adjust the Metered Gas Needle Valve located on the top of the Quest Reactor Unit
until you see a steady stream of gas bubbling through the oil bubbler when one of the
Luer Plugs is removed from the Upper Manifold. This will ensure the establishment
of an inert atmosphere when removing the Luer Plugs to add solvent or reagent.
Modify the gas flow
by
turning the Metered Gas Needle Valve counter clockwise
(left) to increase the gas flow rate and clockwise (right) to decrease the gas flow rate.
Add anhydrous NMP
(-4
mL) to the RVs. Remove the Luer Plug from the first RV
Luer Port (twist and pull) and insert the syringe needle below the Teflon lines of the
male Teflon RV fittings (below the stream of Nz). Take care not to score the interior
surfaces of the Luer Port with the needle. Marring the surface can affect the sealing
of the Luer Port. After delivery, replace the Luer Plug (twist and push) and add NMP
to the remaining
9
RVs in the same manner as the first.
Turn the agitation on by pressing the Agitation ON soft key located to the right of the
Quest Controller Unit display.
Agitate the resin for -5 minutes.
Argonaut Technologies
Part Number 102298
04/02/98
Quest Tutorial