H.
Addition of Reagent Under Nitrogen to the Reaction Vessels
1.
Adjust the Metered Gas Needle Valve located on the top of the Quest Reactor Unit
until you see a steady stream of gas bubbling through the oil bubbler when one of the
Luer Plugs is removed from the Upper Manifold.
2.
Using a syringe add the alcohol or carbonyl diversity reagent in
NMP
to each
RV.
a. Remove the Luer Plug from the first RV Luer Port (twist and pull) and insert the
syringe needle below the Teflon lines of the male Teflon
RV
fitting. Take care
not to score the interior surfaces of the Luer Port with the needle. Marring the
surface can affect the sealing of the Luer Port.
b. Dispense
2
mL of
(1
-2-methoxybenzoy1)-2-pyrrolidinemethanol
in NMP to the
first RV.
c.
Replace the Luer Plug (twist and push) onto the Upper Manifold.
d. Add the appropriate alcohol or carbonyl solution in NMP to the remaining
9
RVs
in the same manner as the first as described in
Table
3.
3.
Using a syringe, add
1
mL of catalyst solution in NMP to each RV in the same
manner as the alcohol and carbonyl diversity reagent (steps 1 and 2a-2d above).
4.
Blanket RVs with nitrogen by turning the
left
Manifold Control Valve to "CLOSED".
Then, turn the
right
Manifold Control Valve to "CLOSED.
Drain Gas Metered Gas
Utility
1
Drain Gas
Utility
2
Metered Gas
Utility 1 Utility
2
5.
Toggle the "Upper Manifold Membrane" switch to "CLOSED". Please refer to the
Quest User Manual for the function of the Upper Manifold Membrane Switch.
I.
Adjusting the Agitation Bar
1.
Raise the Agitation Bar up by pushing the Agitation UP soft key located to the right
of the Quest Controller Unit display.
2.
Use the
7/64"
hex wrench supplied with the Quest to loosen the Agitation Stops.
Adjust the Agitation Stops on the Reactor Unit so the maximum magnet height is
-2
cm below the solvent level in the RVs. Tighten the Agitation Stops so both stops are
at the same height.
Argonaut Technologies
Part
Number
102298
04/02/98
Quest Tutorial