9.
The heaters
will
shut off when the timer runs out and the RVs
will
begin cooling to
ambient temperature.
-----------------
A
(ON)
B
(OFF)
SET
60.0 20.0
ACT
60.9
16-2
02:59:35 0O:OO:OO
Figure
15.
Temperature monitor display
K.
Draining the RVs
1. After the instrument has returned to room temperature (-30 min.), open the Upper
Manifold Membrane by following these steps:
a. Vent the Upper Manifold by turning the Manifold Control Valves to "CLOSED"
and "VENT".
b. Supply Metered Gas to the Upper Manifold
before
opening the Upper Manifold
Membrane by turning the Manifold Control Valves to "CLOSED" and
"METERED GAS".
c. Toggle the Upper Manifold Membrane Switch to "OPEN.
2.
Move the agitator up by pushing the Agitation
UP
soft key located to the right of the
Quest Controller Unit display.
3. Drain all 10
RVs in parallel to the Waste Tank by turning the Manifold Control
Valves to "DRAIN GAS" and "CLOSED". Open the Lower Manifold Drain Valves
by pushing the Drain Levers down. Purge the resin with nitrogen for -1
0
seconds.
4.
Close the Lower Manifold Drain Valves by flipping the Drain Levers upward.
L.
Washing the Resin with
NMP
1. Change the
4
L
bottle of DCM to a
4
L bottle of NMP by doing the following:
a. Vent the Upper Manifold by turning the Manifold Control Valves to "CLOSED"
and "VENT". Wait
-5
seconds.
b. Vent the
4
L Solvent Bottle by turning the black valve on the Solvent Bottle Cap
to the "Vent" position. Allow the soIvent bottle line to vent for -1 minute.
c. Turn both Manifold Control Valves to "CLOSED".
d. Unscrew and remove the Solvent Bottle Cap from the
4
L DCM Solvent Bottle.
e. Wipe off any residual solvent from the Solvent Bottle Pick-Up Line.
f. Throw away the white polyethylene SolventIReagent Inlet Filter
and
install a new
one on the end of the Solvent Bottle Cap Pick-Up Line.
g.
Replace the
4
L
solvent bottle in the Safety Container with a
4
L
bottle of NMP.
Argonaut Technologies
Part
Number 102298
04/02/98
Quest Tutorial