On the Quest Controller Unit press the MODE soft key located below the LCD
display until the Agitation menu is displayed. Set parameters as follows:
Mix
Every
4.0 sec and %Upward=
60%.
Turn the agitation on by pushing the Agitation ON soft key located to the right of the
Quest Controller Unit display. If solvent is splashing up onto the Upper Manifold,
readjust the height of the agitation. The Mix Speed Needle Valve located on the front
of the Quest Controller Unit may also be used to soften or slow the agitation by
turning the valve clockwise.
Setting the Reaction Temperature and Time
Select the Set Temperature menu on the Quest Controller Unit by pressing the MODE
soft key located below the LCD display (Figure 14).
-------------
I
--
-
Set Temperature
-
-
A:
60C
03:OO
ON
B
:
20C 00.00
OFF
RV
Size
5mL
5mL
Figure 14. Set Temperature and time menu
Use the Right or Left
(+,)
soft keys to move the cursor to line
A:
(representative of
Side
A,
RVs 1
-
10).
Use the same soft keys to move the cursor across to the temperature parameter. Use
the PARAMETER SETTING
(?$)
soft keys to set the temperature to 60 "C.
Use the Right or Left soft keys
(*,)
to move the cursor to the time setting. Use a
combination of the Right or Left
(',)
and
the PARAMETER SETTING
(T&)
soft
keys to set the time to 03:OO (3 hours).
Use the Right or Left soft keys
(',)
to move the cursor to the RV size settings. Use
the PARAMETER SETTING
(?&)
soft keys to select the
5
mL RV size.
Use the Right or Left
(',)
soft keys to move the cursor to line A. Press the red and
blue STARTISTOP soft key located beneath the Quest LCD display to turn the
heaters on. The cursor must be on line
A
of the Set Temperature menu to activate the
heaters for Side A.
Select the Monitor Temperature display on the Quest Controller Unit by pressing the
MODE soft key (Figure
15).
When the heaters reach the programmed temperature, the timer will begin counting
down fiom 03:00:00 hours.