8.
Move the agitator up by pressing the Agitation UP soft key located to the right
of
the
Quest Controller Unit display.
9.
Turn the Manifold Control Valves to "DRAIN
GAS"
and
"CLOSED".
Drain Gas Metered Gas
Utility 1 Utility
2
10. Use the white Teflon Drain Lever to
OPEN
the Lower Manifold Drain Valves and
drain all 10 RVs in parallel to the Waste Tank. Allow the nitrogen to purge the resin
for
-30
seconds.
1
1. Raise the Teflon Drain Lever
and
close the Lower Manifold Drain Valves.
12. Re-establish a nitrogen flow through the Upper Manifold to the oil bubbler.
a. Vent the Upper Manifold by turning the Manifold Control Valves to "CLOSED"
and
"VENT".
Drain Gas Metered Gas
Utility 1 Utility 2
b. Turn the Manifold Control Valves to "UTILITY
1" and "METERED GAS".
Drain Gas Metered Gas
Utility
1
Utility
2
13.
Repeat steps
5-
10 with anhydrous NMP.
14.
Allow the nitrogen to purge the resin for -1 minute.
15. Raise the white Teflon Drain Lever and close the Lower Manifold Drain Valves.
16.
Remove the white Teflon Lever from the Lower Manifold Drain Valves and set aside
for later.
17. Be sure to Vent the Upper Manifold and re-establish the nitrogen flow across the
Upper Manifold by repeating step
12
above.
Argonaut Technologies
Part
Number 1022%
04/02/98
Quest Tutorial