D.
RV Removal
&
Instrument Cleaning
1. To remove the RVs from Side A (RVs 1-10) use the following protocol.
a. Position the rotating lever on the
right
side of the Reactor Unit so that is sticking
out from the right side of the Quest parallel to the bench top. Rotate the lever in
towards the front of the Quest
2
10
Reactor Unit. Insert the rotating lever into the
slot in the blue handles of the Upper Manifold (early production Quest
2
10
Systems have two steel pins instead of the slot in the Upper Manifold handles).
b.
While pushing the silver button in on the
right
Upper Manifold handle pull the
rotating lever up. Slide the
right
side of the Upper Manifold up as far as possible.
c. Slide the
left
side of the Upper Manifold up using the
left
rotating lever in the
same manor as the right side. This end should lift much easier now that the right
side of the Upper Manifold has been raised.
d. Push the silver buttons in on both Upper Manifold handles and slide the Upper
Manifold up until the silver buttons snap out indicating that the manifold is locked
into position.
2.
Using the RV extraction tool (red pliers) supplied in the Quest starter kit, remove the
old RVs (RVs 1-10, Side
A).
To remove the RV, grasp the top of the RV
(Figure
4)
Figure
4.
Grasp the top
of
the RV with the RV extraction tool
Argonaut Technologies
Part
Number 1022%
04/02/98
Quest Tutorial