System Overview
Reactor Assembly
Manifold Control Valves
Each side of the reactor assembly contains two manifold control valves: Control Valve
1
(on left) and Control Valve
2
(on right).
Figure 2-1 Quest Manifold Control Valves (Solvent Delivery Configuration)
Drain Metered Gas
Solvent
0
Closed Vent
0
Closed
Utility
1
Utility
2
Control Valve 1 can allow either solvent or drain gas (30 psi) into the corresponding
upper manifold common pathway.
Control Valve
2
can be set to Vent, used during solvent addition, or it can be set to
"Metered Gas" which introduces 10 psi gas into the corresponding upper manifold
common pathway. Metered Gas is used primarily to provide an inert atmosphere to the
RVs and to provide a lower pressure drain gas for product collection into vials. The
flow rate of the metered gas can be controlled via a needle valve located at the top of
the Quest unit.
The manifold control valves are set in conjunction with each other to create various
fluidic paths.
Some possible settings are as follows:
Solvent Delivery.-
Creates a flow path to deliver solvent from the
external
4L
bottle, in parallel to the bank of 10 RVs.
Solvent Line Flush
-
Creates a path for gas at 30 psi to flow through
the manifold after a solvent delivery to blow the residual solvent into
the RVs.
Vessel
Draining
-
Supplies
30
psi gas to the upper manifold common
path, pressurizing the upper RV head space for draining of waste
liquid to the glass waste tank.