Operating
the
Quest 210
Starting
a
Reaction
Adding Solvent or Reagents Through the Luer Ports
Solvents, liquid reagents and solids can be added to the reaction vessels through the luer
ports located at the top of the upper manifold. Typically, these reagents are added via
syringe or pipette.
1.
Prior to adding solvents or reagents through the luer ports, it is recommended that
the system be vented, as it may be under pressure. To vent the vessels, rotate the
Manifold Control Valves as shown below. Slowly approach the "Vent" setting on
the valve to vent the pressure slowly to avoiding "bumping" the solution.
drain
solvent closed
utility
1
metered gas
vent
0
closed
utility
2
-
2.
After placing the Manifold Control Valve to the "Vent" position, slowly remove the
her plug by rotating the plug as it is pulled back.
WARNING The reaction vessel contents may be under pressure. Wear safety glasses,
gloves and appropriate clothing.
-
Do not open the upper luer ports when the RVs are being heated. Turn off the heaters
and allow the vessels to return to ambient temperature.
-
3.
To add liquid reagents or solvents, use a syringe with needle, or pipette.
-
WARNING Be careful not to score the interior surfaces of the her port with the
needle, as marring the surface can affect the sealing of the her port.
4.
For solids, use a spatula or weigh paper to add the solids. If residual solid is present
on the interior luer bore surface, rinse the solid down with a small amount of
solvent.