System
Overview
Controller Unit
LCD
Screen and Keypad
The Controller Unit
LCD
screen and keypad allow you to set the reaction temperature,
temperature duration, and agitation parameters.
LCD
Display
The
LCD
display presents the various Quest programming screens. These screens allow
you to change reaction temperature, temperature duration, and agitation parameters.
Also, the screens will display status on the agitation and temperature settings.
Use the key pad buttons to move between screens, select a parameter on the active
screen, and increase or decrease the value of the chosen parameter. For more
information on the Mode, Forward and Back, Parameter,
STARTISTOP, and Agitator
ON control box buttons, see the definitions provided in this section.
The possible display screens are as follows:
Sign-on Screen
-.identifies this instrument as Argonaut
Technologies' Quest 2 10
Firmware Version Screen
-
shows the Quest 210 firmware version
number
Agitation Control Screen
-
allows you to set the agitation parameters.
You can set three agitation parameters on this screen:
MixEveryseconds, Upstroke, and
%
Upward parameters.
The MixEvery-seconds selection is where the agitation frequency is set (the time
between consecutive up-strokes.)
The Upstroke parameter is the amount of time the agitation bar will be in the upper
position (This number is necessarily less than the agitation frequency value). This time
begins with the Upstroke initiation and ends with downstroke initiation.
The
%
Upward parameter controls the upstroke in the same manner as the "Upstroke"
parameter, but is expressed in terms of a percentage of the agitation frequency, rather
than a time.
Set Temperature Screen
-
allows you to set a temperature for each of
the 2 reaction vessel banks, turn on the heaters, designate how long
each reaction vessel bank should remain at this temperature, and also
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com