Operating the Quest
21
0
Starting a Reaction
To stop the gas flow set the Manifold Control Valves as follows:
drain
solvent closed
utility
1
metered
gas
vent
0
closed
utility
2
Collection of Liquid Reaction Products
When it is time to collect reaction products from the Quest, the waste container can be
removed and an appropriate collection rack installed. Position the collection rack on the
waste tray, raise the rack up until the tubes on the Lower Manifold Drain Block are
below the lip of the vials, and lock in place using the buttons on the handles.
NOTE For product collection, use the
10
psi Metered Gas, as opposed to the
30
psi
Drain Gas to provide the head pressure for fluid transfer. Using the higher pressure can
result in splashing.
Collection Rack Options
Collection racks
are
holders that can contain up to 20 screw top tubes or vials into which
you can collect product. The following racks are available for product collection:
A rack for
13mm
x
100
mm
screw cap tubes
A rack for 20
ml
or 40
rnl
scintillation vials
Draining Synthesis Products to the Collection Vials
-
-
----
-
WARNING
a
During product collection, there can be
a
small amount of splashing.
-
Wear safety glasses, gloves, and appropriate clothing.