Operating the Quest
210
Starting
a
Reaction
Changing
a
Solvent Bottle
Changing the
4L
solvent bottle involves back-flushing the solvent delivery line,
removing the bottle, and changing the end line filter.
1.
Reaction Vessels need to be in place on at least
1
side of the Quest.
2.
Toggle the "Upper Manifold Membrane" switch to "Open."
3.
On the solvent bottle cap, turn the dial to the "Vent" position. Wait
1
minute for the
bottle to vent before proceeding.
4.
For the side of the Quest with the reaction vessels installed, vent the reaction vessels
by
setting the control valves as follows: Toggle the "Upper Manifold Membrane"
switch to "Open."
5.
Next, the reaction vessels will be vented of residual pressure. Set the Manifold
Control Valves as follows:
drain
solvent closed
utility
1
metered gas
vent
0
closed
utility
2